Cardiotocography

Cardiotocography
Intervention

A cardiotocograph recording fetal heart rate and uterine contractions
ICD-9-CM 75.32
MeSH D015148

In medicine (obstetrics), cardiotocography (CTG) is a technical means of recording (-graphy) the fetal heartbeat (cardio-) and the uterine contractions (-toco-) during pregnancy, typically in the third trimester. The machine used to perform the monitoring is called a cardiotocograph, more commonly known as an electronic fetal monitor (EFM).

The invasive fetal monitoring was invented by Doctors Orvan Hess and Edward Hon. A refined (antepartal, non-invasive, beat-to-beat) version (cardiotocograph) was later developed for Hewlett Packard by Dr. Konrad Hammacher.

Contents

Method

Simultaneous recordings are performed by two separate transducers, one for the measurement of the fetal heart rate and a second one for the uterine contractions. Each of the transducers may be either external or internal.

External measurement means taping or strapping the two sensors to the abdominal wall. The heart ultrasonic sensor, similar to a Doppler fetal monitor, detects motion of the fetal heart. The pressure-sensitive contraction transducer, called a tocodynamometer (toco), measures the tension of the maternal abdominal wall - an indirect measure of the intrauterine pressure.

Internal measurement requires a certain degree of cervical dilatation, as it involves inserting a pressure catheter into the uterine cavity, as well as attaching a scalp electrode to the fetal head to adequately measure the electric activity of the fetal heart. Internal measurement is more precise, and might be preferable when a complicated childbirth is expected.

A typical CTG reading is printed on paper and/or stored on a computer for later reference. Use of CTG and a computer network, allows continual remote surveillance: a single obstetrical nurse, midwife, or obstetrician can watch the CTG traces of multiple patients simultaneously, via a computer station.

Interpretation

In the US, the Eunice Kennedy Shriver National Institute of Child Health and Human Development sponsored a workshop to develop a standardized nomenclature for use in interpreting intrapartum fetal heart rate and uterine contraction patterns. This nomenclature has been adopted by the Association of Women’s Health, Obstetric, and Neonatal Nurses (AWHONN), the American College of Obstetricians and Gynecologists (ACOG), and the Society for Maternal-Fetal Medicine. [1]

The Royal College of Obstetricians and Gynaecologists[2] and the Society of Obstetricians and Gynaecologists of Canada [3]have also published consensus statements on standardized nomenclature for fetal heart rate patterns.

Interpretation of a CTG tracing requires both qualitative and quantitative description of:

Uterine activity

There are several factors used in assessing uterine activity.

The NICHD nomenclature[1] defines uterine activity by quantifying the number of contractions present in a 10-minute window, averaged over 30 minutes. Uterine activity may be defined as:

Baseline fetal heart rate

The NICHD nomenclature[1] defines baseline fetal heart rate as: The baseline FHR is determined by approximating the mean FHR rounded to increments of 5 beats per minute (bpm) during a 10-minute window, excluding accelerations and decelerations and periods of marked FHR variability (greater than 25 bpm). There must be at least 2 minutes of identifiable baseline segments (not necessarily contiguous) in any 10-minute window, or the baseline for that period is indeterminate. In such cases, it may be necessary to refer to the previous 10-minute window for determination of the baseline. Abnormal baseline is termed bradycardia when the baseline FHR is less than 110 bpm; it is termed tachycardia when the baseline FHR is greater than160 bpm.

Baseline FHR variability

The NICHD nomenclature[1] defines baseline FHR variability as: Baseline FHR variability is determined in a 10- minute window, excluding accelerations and decelerations. Baseline FHR variability is defined as fluctuations in the baseline FHR that are irregular in amplitude and frequency. The fluctuations are visually quantitated as the amplitude of the peak- to-trough in bpm. Using this definition, the baseline FHR variability is categorized by the quantitated amplitude as:

Accelerations

The NICHD nomenclature[1] defines an acceleration as a visually apparent abrupt increase in FHR. An abrupt increase is defined as an increase from the onset of acceleration to the peak in less than or equal to 30 seconds. To be called an acceleration, the peak must be greater than or equal to 15 bpm, and the acceleration must last greater than or equal to 15 seconds from the onset to return. A prolonged acceleration is greater than or equal to 2 minutes but less than 10 minutes in duration. An acceleration lasting greater than or equal to 10 minutes is defined as a baseline change. Before 32 weeks of gestation, accelerations are defined as having a peak greater than or equal to 10 bpm and a duration of greater than or equal to 10 seconds.

Periodic or episodic decelerations

Periodic refers to decelerations that are associated with contractions; episodic refers to those not associated with contractions. There are four types of decelerations as defined by the NICHD nomenclature.[1]

Additionally decelerations can be recurrent or intermittent based on their frequency (more or less than 50% of the time) within a 20 min window.[1]

FHR pattern classification

The NICHD workgroup proposed terminology of a three-tiered system to replace the older undefined terms "reassuring" and "nonreassuring".[1]

Effect on management

A Cochrane Collaboration review has shown that use of cardiotocography reduces the rate of seizures in the newborn, but there is no clear benefit in the prevention of cerebral palsy, perinatal death and other complications of labour. In contrast, labour monitored by CTG is slightly more likely to result in instrumental delivery (forceps or vacuum extraction) or Cesarean section.[4] The false-positive rate of cardiotocography for cerebral palsy is given as high as 99%, meaning that only 1-2 of one hundred babies with non-reassuring patterns will develop cerebral palsy.[5] The introduction of additional methods of intrapartum assessment has given mixed results.[6]

When introduced, this practice was expected to reduce the incidence of fetal demise in labor and make for a reduction in cerebral palsy (CP). Its use became almost universal for hospital births in the U.S. In recent years there has been some controversy as to the utility of the cardiotocograph in low-risk pregnancies, and the related belief that over-reliance on the test has led to increased misdiagnoses of fetal distress and hence increased (and possibly unnecessary) cesarean deliveries.[7]

Manufacturer

The popular manufacturers are GE (corometrics), Philips, Huntleigh (Sonicaid), Analogic, Ultrasound Technologies (Seward/Wakeling), Toitu, and Sunray.

References

  1. ^ a b c d e f g h Macones GA, Hankins GD, Spong CY, et al.. "The 2008 National Institute of Child Health and Human Development workshop report on electronic fetal monitoring:update on definitions, interpretation, and research guidelines.". Obstet Gynecol (2008) 112:661-666. 
  2. ^ NICE Guideline Intrapartum care: management and delivery of care to women in labour
  3. ^ SOGC Fetal health Surveillance: antepartum and intrapartum Consensus Guideline
  4. ^ Alfirevic, Zarko; Devane, Declan; Gyte, Gillian ML (2006). Alfirevic, Zarko. ed. "Continuous cardiotocography (CTG) as a form of electronic fetal monitoring (EFM) for fetal assessment during labour". Cochrane Database of Systematic Reviews. doi:10.1002/14651858.CD006066. 
  5. ^ American College of Obstetricians and Gynecologists (2005). "ACOG Practice Bulletin. Clinical Management Guidelines for Obstetrician-Gynecologists, Number 70, December 2005 (Replaces Practice Bulletin Number 62, May 2005). Intrapartum fetal heart rate monitoring". Obstetrics and gynecology 106 (6): 1453–60. PMID 16319279. 
  6. ^ Pettker, Christian M.; Lockwood, Charles J. (2008). "Opinion: New FHR monitoring standards: Something old and something new". Contemporay OB/GYN. http://www.modernmedicine.com/modernmedicine/Modern+Medicine+Now/Opinion-New-FHR-monitoring-standards-Something-old/ArticleStandard/Article/detail/571513. 
  7. ^ Goddard, R. (2001). "Electronic fetal monitoring". BMJ 322 (7300): 1436–1437. doi:10.1136/bmj.322.7300.1436. PMC 1120506. PMID 11408285. http://www.pubmedcentral.nih.gov/articlerender.fcgi?tool=pmcentrez&artid=1120506. 

See also